IR9. Misleading Peaks
There are some practical problems that can make IR
interpretation in real life more difficult. Being aware of these problems may
make you double-check your suspicions:
- water in the sample. Since water contains O-H bonds, water in a sample
will make it appear as if the compound contains O-H bonds. There are
experimental techniques for removing water from a sample but they must be
done carefully.
- overtones in a spectrum. Overtones are absorptions occurring at
different multiples of the normal frequency. Strong overtones of carbonyl
peaks often occur at about twice the normal wavenumber. A large peak at 1750
cm-1 might be accompanied by a smaller peak at 3500 cm
-1,
and could be confused with an O-H peak.
In addition, there are complications that you may run into based on the
instrument or technique used to obtain the spectrum.
- Note that the x-axis scale on the spectrum is not uniform: the scale is
indexed every 1000 cm
-1
between 4000 and 2000
cm-1 and every 500
cm-1
below that. Spectra from different
instruments may display this scale differently.
Different techniques may lead to increased absorption at one end of the
spectrum vs. the other. For example, an OH stretch is much more
dominant if the sample is prepared in a KBr mull than if the spectrum is
obtained via attenuated total reflectance.
Back Next
This site is written and maintained by Chris P. Schaller, Ph.D., College of Saint Benedict / Saint John's
University (with contributions from other authors as noted). It is freely
available for educational use.
Structure & Reactivity in Organic, Biological and Inorganic Chemistry by Chris Schaller is licensed under a Creative Commons Attribution-NonCommercial 3.0 Unported License.
Send corrections to cschaller@csbsju.edu
Navigation:
Back to Infrared Spectroscopy
Back to Structure Determination
Back to Structure & Reactivity