Biology 324 -- Spring 2001

COMPOUND ACTION POTENTIALS: FROG SCIATIC NERVE

[Adapted from Oakley, B. and R. Schafer.  1978.  Experimental neurobiology.  Univ. Michigan, Ann Arbor.]

  Introduction

In 1850, Hermann von Helmholtz first estimated the velocity of nerve impulses transmitted in a frog nerve-muscle preparation with a mechanical kymograph and writing levers.  During the latter half of the 19th century, concepts underlying the modern theory of nerve conduction were developed by Sherrington and others, but modern electro-physiological research awaited the development of the cathode ray oscilloscope.  Using this apparatus, Erlanger and Gasser in 1921 first measured the ionic currents of compound action potentials. Their studies provided an important foundation for our present understanding of nerve function.  The frog sciatic nerve was the classical preparation for study of the action potential until experimental researchers developed intracellular recording methods for studying squid giant fibers.

Action potentials can be elicited simultaneously in thousands of axons of a peripheral nerve like the sciatic by electrical stimulation.  The collective response is termed a compound action potential.  Indiscriminate as this gross recording technique may seem, some basic aspects of neuronal conduction--maximal firing rates, threshold, conduction velocity, and the role of axon size and myelination--can be demonstrated using the whole nerve approach.

Procedure

Preliminary setup:  It is essential to familiarize yourself with the instruments and recording system before beginning the dissection.  The instrumentation may seem formidable, but generally it is the faltering viability of a biological preparation that ends the experiment. Therefore, any practical or conceptual problems regarding the equipment should be cleared up before the experimental animal is touched.

Surgical procedure:  You will be provided with a doubly-pithed Bullfrog (Rana catesbiana). Pick up a fold of skin at midabdomen with forceps and, avoiding cutting into the abdominal cavity with your scissors, cut the skin all the way around the frog.  Pull the skin down, everting it as you pull, and strip it off the legs.  Lay the frog dorsal side up and gently separate the thigh muscles to reveal the white sciatic nerve and accompanying blood vessels (See Fig. 1).  Don't mistake white muscle fascia for the nerve.  Ask if you're not sure.

Separate the muscles and free the nerve from surrounding tissue using blunt glass tools whenever you must touch the nerve.  Make such a tool by heating a glass rod in a bunsen burner flame and pulling out a working tip smoothed to about the width of a dulled pencil lead.  Apply amphibian perfusion fluid (Frog Ringer's solution) liberally as you work.   Do not contaminate the nerve with cut tissues or blood, and do not touch the nerve with metal tools or with your fingers. Avoid stretching, pinching, or drying out the nerve.

Hold the urostyle up, and carefully cut the muscles on both sides of the bone.  Free the caudal end of the urostyle and lift it up to expose the underlying structures.  Note the two regions of white fibers that compose the sciatic nerve plexus.  Each sciatic nerve originates as three spinal nerve roots.  Cut the urostyle at its hinge.  Carefully tie the roots together with the end of a 10 cm length of Ringer's-soaked cotton thread.  Cut the nerve roots as close to the spinal cord as possible.  Now free the nerve from the hip to the knee, lifting with the thread as necessary.  [CAUTION: Do not stretch the nerve!]  When the nerve has been totally freed, cut through the distal end with scissors.  Immerse the nerve in a small beaker of perfusion fluid.

Installing the nerve in the recording apparatus:  Plug any holes in the nerve chamber with Vaseline, and fill the chamber with perfusion fluid to a point about 5 mm above the electrode wires.  Lay the nerve lengthwise in the chamber so that it floats above the wires.  Note which end of the nerve is which (anterior end is thicker). Manipulate the nerve with glass tools as you draw off enough of the fluid so that it comes to rest on the electrode wires.  The nerve must be in physical contact with each of the wires, and the level of the fluid must be well below all of the wires to prevent them shorting out.  One end of the nerve may remain in the fluid, but not both.  Place the cover over the nerve chamber to prevent drying.

If drying of the nerve tissue seems to be a problem, add a layer of mineral oil saturated with Ringer's atop the fluid already in the chamber.  Cover the electrodes and the nerve.  As oil is added it may lift the nerve off the electrodes.  To prevent this, add the oil/Ringer's mixture by dropping it over and on top of the nerve until the nerve is immersed.  Look to see that good contact is made between the nerve and each electrode.  If contact between the nerve and electrode wires is lost, it can be reestablished by manipulating the nerve using a dropper and a half-squeezed-out drop of Ringer's solution.

Analog recording procedure:  Arrange the electrode leads so that you stimulate and record at opposite ends of the nerve and ground the center (See Fig. 2).  For recording, connect a pair of cables to two electrodes near the distal (thin) portion of the nerve and connect the other end of this pair of cables to the input of the preamplifier [or direct to the oscilloscope if preamplifiers are not used].  These cables should be as short as possible to minimize the pickup of electrical interference.

Connect a third wire (green if possible) to one of the other electrodes about midway along the nerve and run it to a ground terminal on the preamplifier.  Connect another pair of cables from the stimulator output to a pair of electrodes at the proximal (thick) end of the nerve.  Make sure that the negative electrode is nearest the recording electrodes.  The action potential is initiated at the negative electrode (cathode).

The presence of the anode (positive electrode) between the cathode and recording electrodes may block AP transmission since the anode hyperpolarizes the nerve.

Connect the output of the preamplifier to the input of the oscilloscope with appropriate shielded cables and connectors.  Connect the stimulator sync out (trigger output) to the trigger input of the oscilloscope.  This arrangement synchronizes the initiation of the oscilloscope sweep with the output pulse of the stimulator.  Refer to Fig. 3 for the recording setup.

Use the following initial settings on your equipment:

Stimulator Preamplifier
Frequency 7/sec Gain 100X
Duration 0.1 msec Low band pass filters = 10 Hz
Voltage 0.1 V to start Low
Mode off at first High band pass filters = 3-5 kHz
  Input on USE

 

Oscilloscope
Time Base: Vertical Amplifier:
Time/division = 1 msec/div Volts/div = 20 mV/div (check that VAR knob is set fully clockwise to CAL)
Triggering mode = External/Normal Position = Trace at midscale or below
Input selector = DC mode

Calibration:  Adjust the overall system gain (preamplifier plus oscilloscope) to about 100 mV/div.  Check by using the preamplifier calibration function.  Set the Grass preamp input knob to CAL 100 mV and depress the G1 NEG button several times in succession.  Alter the oscilloscope vertical amplifier gain so as to produce a 1 cm deflection when G1 is pulled NEG.  During the experiments, you may need to alter the system gain to best display the compound AP's that the nerve produces, and you should recalibrate using this approach when doing so.

Recall that these are suggested settings for starting the experiment.  Readjusting the gain of the vertical amplifier and the time base of the oscilloscope to visually display the nerve action potential is an ongoing process.  Just keep accurate notes on the settings used whenever you record a piece of data. 

Digital recording with MacLab:  Turn the MacLab and Macintosh on.  Open the folder for your lab group by double-clicking on the icon.  This folder should contain all the software you will need to run and analyze today's lab.  Run the program called SCOPE by double-clicking on the icon labeled "Sciatic Nerve Lab".  This will launch SCOPE and provide you with a ready-to-record computerized oscilloscope.  This digital oscilloscope differs from the Kikusui in several ways, but most important for us is the MacLab's ability to record and store a waveform for analysis.

Plug in the Kikusui oscilloscope's CH 1 output (on the back of the machine) to the MacLab's input CH 1.  With the Kikusui on and free-running (trigger = auto) but the stimulator MODE control off and the nerve quiescent, open the MacLab input amplifier dialog box in SCOPE (just point and click with the cursor).

With the Grass preamplifier input knob on CAL 100 mV, hit the G1 NEG button on the preamp several times.  This should produce a 1 cm deflection on the oscilloscope (since you've already calibrated the overall system gain up to that point) and a also should produce a good sized square wave ("good-sized" being about 1/3 full scale or so) on the SCOPE input amplifier recording trace.  Reset the oscilloscope vertical amplifier gain or the SCOPE input amplifier gain (click and drag with the cursor) to give an easily visible wave on the computer when G1 NEG is pushed.  This CAL value from the Grass preamp can be used later to calibrate the computer recordings, so once you've started, record a CAL wave or two of known size to go along with your recorded nerve action potentials.  When you're satisfied, close the SCOPE input amplifier box by clicking on OK.

Plug the stimulator trigger output into the MacLab trigger input using the “pulse stretcher” box.  Check the display dialog box in SCOPE to examine the trigger settings under recording.  This should be set for external.  Note the recording settings too--multiple gets really busy really fast, so you probably should use single sweeps or overlay mode at first to record.

When you're ready to record a wave from your nerve preparation (later, not yet!), you will set up the oscilloscope to sweep and display the AP.  When SCOPE is triggered, manually with the mouse (USER), or by the stimulator (TRIGGER), a wave will appear on the screen.  You can record the displayed wave or choose New Data to record another one.  Practice using the SCOPE recording feature without recording AP's until you've got it down.  Once recording, note your stimulus voltage and other data on the comments notebook attached to each scope “page”.  Record a calibration pulse to use in measuring the size of the AP's.  See the SCOPE instruction manual for more detail.

Experimental procedure

Now that you have reached a good understanding of the equipment's setup and operation, you're finally ready to start the experiments.  Read through each section in advance and know what the goal of that experiment is before you begin.

1.  Threshold:  First, activate the stimulator by placing the output mode switch in the continuous (multiple) position.  Gradually increase the stimulus voltage from 0.1 V.  You will see the stimulus artifact as the first-appearing wave; this is the stimulating voltage conducted on the outside of the nerve and picked up through the recording electrodes.  The artifact can be seen to vary with stimulus duration.

Continue to increase the stimulus voltage until a second wave appears to the right of the artifact.  This is the compound action potential.  Continue to increase the stimulus until this wave reaches a maximum amplitude.  Reduce the voltage and note the voltage at which the AP first appears.  This is the threshold voltage for the most sensitive axons (or those most accessible to the stimulating current).

Increase stimulus intensity until a maximal response is seen.  At this point all the nerve fibers are actively conducting AP's and the waveform seen is the sum of all of them.  This growth of the AP with increasing stimulus intensity obscures the fact that the action potential of each individual fiber is an all-or-none event.  The compound AP has these distinguishing properties:  It is not the first deflection observed; its amplitude, though initially increased by raising stimulus intensity, is not a linear function of stimulus strength; its duration is not a direct function of stimulus duration; it does not have the shape of the stimulus artifact.

Record threshold and the voltage needed to recruit a maximal response.  Record a typical waveform with Scope in MacLab.  Write down all instrument settings and check the timebase and vertical calibration for your recording.  Turn off the stimulator to allow the nerve to rest.

2. Recruitment of nerve fibers:  To produce a graphical illustration of the response of your nerve to different stimulus intensities, record several waves at differing stimulus voltages between threshold and maximal voltage.

3. Waveform--monophasic and biphasic:  The shape of the waveform observed on the oscilloscope screen depends on a number of factors.  The distance between recording electrodes, sweep rate, gain, filter settings, and condition of the nerve all influence the shape of the compound AP observed.

Return stimulus voltage to 0.1 V.  Turn up the intensity until the volt-age is about 10% above that needed to elicit a maximal response. Reverse the polarity of the recording electrodes, if necessary, so that the initial deflection of the displayed waveform is upward.  The compound AP from an undamaged nerve is usually biphasic.  As the AP sweeps by the first recording electrode, it drives that electrode negative with respect to the more distant electrode.  If you previously arranged the electrodes as above, the initial deflection will be upward.  Then, as the wave of depolarization (the AP) arrives at the second recording electrode, making it negative, the oscilloscope trace is deflected downward.  Record the biphasic wave in Scope, recording amplification and timebase for reference.

  Crush the nerve at the site of the second (more distant) recording electrode by pinching it with a pair of fine forceps.  This should inactivate the nerve and produce a monophasic recording.  You may need to crush and check the nerve several times to ensure that you have a fully monophasic recording.  A tiny drop of isotonic KCl (0.16 M) applied to the crushed area will aid in development of a monophasic AP.  Record the monophasic wave and record amplification and timebase.

4. Conduction velocity:  Measuring the time and distance between appearance of the AP at different recording electrodes can provide an estimate of the speed of nerve AP conduction.  Rearrange electrodes so that you stimulate at the distal (thin) end and record at the proximal (thick) end.  Determine conduction velocity by moving the active (first) recording electrode and recording time and distances as in Fig. 4.  Measure times from the start of the stimulus artifact (or beginning of the sweep) to the peak of the AP.  You can read the oscilloscope display most accurately if you spread it out with a fast sweep speed.  Measure distance traveled using a micrometer.  You can do this after you've finished if you're certain to record the electode numbers used.  Express conduction velocity in meters/second.

If the nerve is very short, the conduction velocity may be estimated by measuring the time interval between the beginning of the stimulus and the distance between stimulating cathode and the first recording electrode.  This measure is less accurate because it includes an unknown time to initiate the impulse.   The accuracy of this method is increased by using a supramaximal stimulus intensity and as brief a stimulus duration as is feasible.

5. Fiber groups:  Within the total population of fibers in frog sciatic nerve there are several groups of axons of similar diameter and therefore, similar threshold and conduction velocity.  Connect the electrodes for monophasic recording.  Reduce the frequency to 5 stimulations/ sec.  Try to identify as many peaks of the compound AP as possible, by slowly increasing the stimulus voltage and looking for the addition of new peaks.  It should be possible to find two of the three major peaks of the compound AP demonstrated by Erlanger and Gasser (1968):  A, the largest, corresponding to large myelinated fibers; and C (the slowest wave), corresponding to very fine unmyelinated fibers.  Within the A wave you may be able to separate several subpeaks, the A-alph, A-beta, and A-delta fibers (see Fig. 5).

The conduction velocity of the C fibers is only 1/100 that of the A-alpha peak and to see the C peak you must therefore stimulate at a low enough rate for it to appear before the next A wave.  The sweep rate should also be low (ca. 50 msec/div) and the stimulus intensity high.

Determine the relative amplitudes, thresholds, and conduction velocities of each group of fibers in your preparation.  Fiber diameter is probably the most important determinant of  conduction velocity, with large fibers conducting faster.

6. Strength-duration curve [optional]:  The ability of the stimulus to elicit a response is dependent on the stimulus duration as well as its intensity.  In other words, a response can be obtained using strong current for a short time or a weak current for a long time.  The relationship between strength and duration can be determined empirically for your sciatic nerve preparation.

Vary the duration and measure the threshold voltage.  You may define threshold as a small but observable response (for example, a 1 cm deflection).  Use a constant criterion for recording the threshold stimulus.  Start by setting the stimulus duration to 100 msec and gradually increase the stimulus intensity until a response in noted.  Decrease the duration to 50 msec and advance the voltage until an identical response is seen.  Continue this process for a number of different stimulus durations.

Plot a strength duration curve, with stimulus intensity (V) on the ordinate and duration (msec) on the abscissa.  Your curve should look approximately like Fig. 6.  The minimum intensity which elicits a response at infinite duration is called the rheobase.  Chronaxie (2X rheobase) is a measure of the excitability of nervous tissue.  The smaller its value, the more excitable the nerve.  These concepts have lost some of the importance they once had for understanding nerve function, but chronaxie is still useful to compare excitability of nerve and muscle tissues.  Strength-duration curves have also been used experimentally to follow the course of nerve and muscle regeneration.

Analysis and report -- to be elaborated in another handout

Record and tabulate values for threshold, voltage for maximal response, conduction velocity.  Estimate conduction velocity and fiber diameter for different fiber groups.  Compare compound and single-cell AP's and mono- and biphasic waves.  Include your digital recordings of observed AP's, with time and vertical scales identified.

Graph response (peak height or mV) vs. stimulus intensity in part 1.  Graph stimulus intensity at threshold or a fixed response vs. duration and determine chronaxie and time constant, if you obtained those data in the optional SD curve experiment.

In your report, discuss the main features of nerve action potentials, including the ionic basis of the AP wave and its propagation.  Explain how conduction velocity varies in different animals [see Prosser 1973; Schmidt-Nielsen 1978; Bullock, Orkand, and Grinnell 1978, all in the lab].

References  [Locate other more recent references by doing searches on your own -- Medline, Medscape, Infotrack (from PhysioLink site)]

Aidley, D.J. 1991. The physiology of excitable cells. 3rd ed. Univ. Press, Cambridge.

Baker, P.F. 1966. The nerve axon. Sci. Am. 214:74-82.

Cragg, B.G. and P.K. Thomas. 1957. The relationships between conduction velocity and the diameter and internodal length of peripheral nerve fibers. J. Physiol. 136:606-614.

Erlanger, J. and H.S. Gasser. 1930. The action potential in fibers of slow conduction in spinal roots and somatic nerves.  Am. J. Physiol. 92:43-82.

Erlanger, J. and H.S. Gasser. 1968. Electrical signs of nervous activity. 2nd ed. Univ. Pennsylvania Press, Philadelphia.

Erlanger, J., H.S. Gasser, and G.H. Bishop. 1924. The compound nature of the action current of nerve as disclosed by the cathode ray oscillograph.  Am. J. Physiol. 70:624-666.

Hill, A.V. 1936. The strength-duration relation for electric excitation of medullated nerve. Proc. Roy. Soc. B. 119:440-453.

Oakley, B. and R. Schafer. 1978. Experimental neurobiology: a laboratory manual. Univ. Michigan Press, Ann Arbor.

Stevens, C.F. 1979. The neuron. Sci. Am. 241:54-65.